Lumax Technical Manual 195
Parameter Range Standard
Rate hysteresis
(Lumax 700/740 &
600/640 only)
OFF -
AV delay 50..(10)..350 ms (fixed
AV delay)
140 ms
Post-shock duration OFF, 00:10 .. (00:10) ..
00:50, 01:00 .. (01:00) ..
10:00 mm:ss
00:10 mm:ss
Post-shock duration
(Lumax 700/740 &
600/640 only)
OFF, 00:10, 00:30, 01:00,
02:00, 05:00, 10:00
mm:ss
00:10 mm:ss
CRT Therapy Parameters
Initially paced
chamber
LV, RV LV
VV delay after Vp 0… (5)… 100 ms 0 ms
Home Monitoring
Home Monitoring OFF, ON OFF
Transmission time Time (hh:mm) [01:00
hh:mm]
Transmission time
(Lumax 700/740 &
600/640 only)
Std., 00:00 ..(01:00) ..
23:00 (hh:mm)
[Std.]
IEGM for therapy
episode
OFF, ON [OFF]
IEGM for therapy
episode
(Lumax 700/740 &
600/640 only)
OFF, ON [ON]
IEGM for monitoring
episode
OFF, ON [OFF]
IEGM for monitoring
episode
(Lumax 700/740 &
600/640 only)
OFF, ON [ON]
Periodic IEGM OFF, 2, 3, 4, 6 months [OFF]
Periodic IEGM
(Lumax 700/740 &
600/640 only)
OFF, Date, 30, 60, 90,
120, 180 days
[90]
Sustained atrial
episode
OFF, 0,5, 6, 12, 18 h [12 h]