194 Lumax Technical Manual
Parameter Range Standard
S3-S4 interval OFF, 200...(10)...600 ms
simply alterable via [+ / -]
buttons
OFF
Minimum ATP
interval
200 ms
Atrial NIPS
Lumax 700/740 & 600/640 Only
Backup Stimulation
(VVI-Rate)
OFF, 30 .. (10) .. 120
bpm
OFF
Report OFF, 5, 10, 25, 50 mm/s Preference
S1-S1 interval 80…(10)…2000 ms 600 ms
S1 cycles 0...(1)...10 7
S1-S2 interval None, 80...(10)...1000 ms
simply alterable via [+ / -]
buttons
None
S2-S3 interval None, 80...(10)...1000 ms
simply alterable via [+ / -]
buttons
None
S3-S4 interval None, 80...(10)...1000 ms
simply alterable via [+ / -]
buttons
None
Burst rate Min. 30…(10)…250 bpm 150 bpm
Burst rate Start 30…(10)…800 bpm 250 bpm
Burst rate Max. 30…(10)…800 bpm 350 bpm
Post-Shock Pacing
Mode DDI at permanent
DDD(R), DDI (R), AAI(R)
VDI at permanent
VDD(R), VDI(R),
VVI at permanent VVI(R),
OFF
Modes (VR-T DX
Only)
AAI
VDI at permanent
VDD(R), VDI(R),
VVI at permanent VVI(R),
OFF
Basic rate 30 .. (5)..100..(10) .. 160
bpm
60 bpm
Rate hysteresis OFF, -5 .. (-5) .. –65 bpm OFF