Lumax Technical Manual 193
Parameter Range Standard
ATP type
(Lumax 700/740 &
600/640 only)
OFF, Burst, Ramp Burst
S1 number 1...(1)...10 [5]
S1 number
(Lumax 700/740 &
600/640 only)
1...(1)...10 [8]
R-S1 interval 200...(10)...350 ms
(absolute);
70...(5)...95 % (adaptive)
[80%]
R-S1 interval
(Lumax 700/740 &
600/640 only)
70...(5)...95 % (adaptive) [85%]
S1 decrement 5...(5)...40 ms [10 ms]
S1-S2 interval 200...(10)...350 ms
(absolute);
70...(5)...95 % (adaptive)
[70%]
Stability 12% 12%
ATP attempts 1 [1]
ATP pulse amplitude 7.5 V 7.5 V
ATP pulse width 1.5 ms 1.5 ms
Ventricular NIPS
Lumax 700/740 & 600/640 Only
ATP type Burst, Ramp, Burst +
PES
Ramp
Attempts 1 1
Ventricular Pacing BiV, LV, RV RV
S1 number 1...(1)...25 [5]
R-S1 interval 200...(10)...600 ms
(absolute);
70...(5)...95 % (adaptive)
[80%]
S1 decrement 5...(5)...40 ms [10 ms]
S1-S2 interval 200...(10)...600 ms
simply alterable via [+ / -]
buttons
290
S2-S3 interval OFF, 200...(10)...600 ms
simply alterable via [+ / -]
buttons
OFF