User Manual

Table Of Contents
192 Lumax Technical Manual
Parameter Range Standard
Polarity (per zone) Normal,
Reverse,
Alternating
Normal
ATP Parameters
ATP type Burst, Ramp, Burst +
PES
[Burst]
ATP type
(Lumax 700/740 &
600/640 only)
Burst, Ramp [Burst]
ATP attempts OFF, 1…(1)...10 3
ATP attempts
(Lumax 700/740 &
600/640 only)
OFF, 1…(1)...10 [OFF]
S1 number 1...(1)...10 [5]
Add. S1 OFF, ON [ON]
R1-S1 interval 200...(10)...500 ms
(absolute);
70...(5)...95 % (adaptive)
[80 %]
R1-S1 interval
(Lumax 700/740 &
600/640 only)
70...(5)...95 % (adaptive) [80 %]
S1 (RAMP)
decrement
5...(5)...40 ms [10 ms]
Scan decrement OFF, 5...(5)...40 ms [OFF]
S1-S2 interval 200...(10)...500 ms
(absolute);
70...(5)...95 % (adaptive)
[70 %]
Minimal ATP interval 200...(5)...300 ms [200 ms]
Minimal ATP interval
(Lumax 700/740 &
600/640 only)
200 ms -
ATP timeout (not
applicable in Lumax
700/740 & 600/640)
OFF,
00:15...(00:15)...05:00
mm:ss
OFF
ATP optimization OFF, ON OFF
ATP pulse amplitude 7.5 V 7.5 V
ATP pulse width 1.5 ms 1.5 ms
ATP One-Shot Parameter (ATP in VF)
ATP type OFF, Burst, Ramp, Burst
+ PES
OFF