Parameter Range Standard
Energy 1st shock
and 2nd shock
VT1, VT2 (model
340/540)
1
OFF; 1..(1)..16..(2)..40 J 40 J
Energy 1st shock of
VT1, VT2
(Lumax 740/640
only)
OFF; 2..(2)..20..(5)..40 J 40 J
Energy 2nd shock of
VT1, VT2
(Lumax 740/640
only)
OFF; 4..(2)..20..(5)..40 J 40 J
Energy 1st shock
and 2nd shock
VF (model 340/540)
1
1..(1)..16..(2)..40 J 40 J
Energy 1st shock of
VF (Lumax 740/640
only)
2..(2)..20..(5)..40 J 40 J
Energy 2nd shock of
VF (Lumax 740/640
only)
4..(2)..20..(5)..40 J 40 J
Number of shocks
(VT1/VT2)
0..(1)..8 [8]
Number of shocks
(VT1/VT2)
(Lumax 700/740 &
600/640 only)
0, 1, 2, 6, 8 [8]
Number of shocks
(VF)
2; 6..(1)..8 8
Number of shocks
(VF)
(Lumax 700/740 &
600/640 only)
6, 8 8
Confirmation (per
zone)
OFF, ON ON
Shock form (per
zone)
Biphasic, Biphasic2 Biphasic