190 Lumax Technical Manual
Parameter Range Standard
Forced termination
(Lumax 700/740 &
600/640 only)
OFF; 1…(1)…10 min [1 min]
Redetection Counter
Redetection counter
VT1
10...(2)...30 [12]
Redetection counter
VT2
10...(2)...30 [10]
SMART Redetection
SMART redetection
(VT1 & VT2)
OFF, ON [ON]
Ventricular Therapy Parameters
Energy 1st shock
and 2nd shock
VT1, VT2 (model
300/500)
1
OFF; 1..(1)..16..(2)..30 J 30 J
Energy 1st shock of
VT1, VT2
(Lumax 700/600
only)
OFF; 2..(2)..20..(5)..30 J 30 J
Energy 2nd shock of
VT1, VT2
(Lumax 700/600
only)
OFF; 4..(2)..20..(5)..30 J 30 J
Energy 1st shock
and 2nd shock
VF (model 300/500)
1
1..(1)..16..(2)..30 J 30 J
Energy 1st shock of
VF (Lumax 700/600
only)
2..(2)..20..(5)..30 J 30 J
Energy 2nd shock of
VF (Lumax 700/600
only)
4..(2)..20..(5)..30 J 30 J
1
Programmed values are stored energies.