User Manual

Table Of Contents
188 Lumax Technical Manual
Parameter Range Standard
Deactivation
criterion Z
3..(1)..8 5
Mode DDI, DDIR at permanent
DDD(R)
VDI, VDIR at permanent
VDD(R)
DDI
[VDI]
Modes (VR-T DX
Only)
VDI, VDIR at permanent
VDD(R)
VDI
Change in basic rate OFF, +5 ... (5) ...+30 bpm +10 bpm
Post-Mode Switch Response (PMSR)
Post-ModeSw rate OFF, +5 ... (5) ...+50 bpm +10 bpm
Post-ModeSw
duration
1..(1)..30 min 1 min
PMT Protection
PMT
detection/terminatio
n
OFF, ON ON
VA criterion 250..(10)..500 ms 350 ms
Detection
Detection/Therapy ON, OFF ON
Interval
Interval AT/AF 100….(10)…250 bpm
(240…..600 ms)
200 bpm
Interval VT1 OFF, 270...(10)...600 ms OFF
Interval VT2 OFF, 270...(10)...500 ms OFF
Interval VF OFF, 200...(10)...400 ms 300 ms
Interval VF
(Lumax 700/740 &
600/640 only)
OFF, 240...(10)...400 ms 300 ms
Detection Counter
Detection counter
VT1
10...(2)...60 [16]
Detection counter
VT2
10...(2)...40 [14]
Detection counter
VF – X
6...(1)...30 8
Detection counter
VF – Y
8...(1)...31 12