Lumax Technical Manual 187
Parameter Range Standard
Sensor gain 1.0; 1.1; 1.3; 1.4; 1.6; 1.8;
2.0; 2.2; 2.6; 3.0; 3.3; 3.7;
4.0; 4.5; 5.0; 6.0; 7.0; 8.0;
9.0; 10; 11; 12; 14; 16;
18; 20; 22; 24; 28; 32; 35;
40
6.0
Sensor gain
(Lumax 700/740 &
600/640 only)
AUTO, Very low (1.3),
Low (3), Medium (6),
High (12), Very high (26)
Medium (6)
Auto Sensor gain OFF, ON OFF
Sensor threshold
(numbers are not
displayed on screen)
Very low = 0
Low = 3
Medium = 7
High = 11
Very high = 15
Medium
Rate increase 0.5; 1..(1)..6 ppm/cycle 2 bpm/cycle
Rate increase
(Lumax 700/740 &
600/640 only)
1, 2, 4, 8 bpm/cycle 2 bpm/cycle
Rate drop 0.25..(0.25)..1.25
ppm/cycle
0.5
ppm/cycle
Rate decrease
(Lumax 700/740 &
600/640 only)
0.1, 0.2, 0.5, 1.0
bpm/cycle
0.5
ppm/cycle
Upper Tracking Rate (UTR)
Upper tracking rate 90..(10)..160 bpm 130 bpm
Upper tracking rate
atrium
OFF, 240 bpm 240 bpm
(N/A
VR-T DX)
Upper tracking rate
atrium
(Lumax 700/740 &
600/640 only)
OFF, 175, 200. 240 bpm 200 ppm
(N/A
VR-T DX)
Mode Switching
Intervention rate OFF, 100..(10)..250 bpm 160 bpm
Intervention rate
(Lumax 700/740 &
600/640 only)
OFF, 120..(10)..200 bpm 160 bpm
Activation criterion X 3..(1)..8 5