184 Lumax Technical Manual
Parameter Range Standard
Modes (Lumax
700/740 & 600/640
models only)
DDD, DDDR, DDI, DDIR,
D00, VDD, VDDR, VDI,
VDIR, VVI, VVIR, V00,
AAI, AAIR, OFF
Modes (VR-T DX
Only)
VDD, VDDR, VDI, VDIR,
VVI, VVIR, OFF
Modes (Lumax
740/640 VR-T DX
Only)
VDD, VDDR, VDI, VDIR,
VVI, VVIR, V00, OFF
VVI
Basic Rate Day/Night
Basic rate 30..(5)..100..(10)..160
bpm
60 bpm
Night rate OFF, 30..(5)..100 bpm OFF
Night beginning 00:00..(1 min)..23:59 h:m [22:00 h:m]
Night ending 00:00..(1 min)..23:59 h:m [06:00 h:m]
Rate Hysteresis
Rate hysteresis OFF, -5..(-5)..-90 ppm OFF
Rate hysteresis
(Lumax 740/640
VR-T DX Only)
OFF, -5…(-5)… -25…
(-20)…-65 bpm
OFF
Repetitive OFF; 1..(1)..15 [OFF]
Scan OFF; 1..(1)..15 [OFF]
Scan / Repetitive
(Lumax 740/640
VR-T DX Only)
OFF; ON (10) [ON(10)]
AV Delay
AV delay Low, medium, high, fixed,
individ.
Low
AV delay 1 40..(5)..350 ms 150 ms
AV delay 1 after
pace
(Lumax 740/640
VR-T DX Only)
15, 40..(5)..350 ms -
AV delay 1 after
sense
(Lumax 740/640
VR-T DX Only)
=AV delay 1 after pace +
Sense compensation
-
AV rate 1 30..(10)..120 bpm 60 bpm