Lumax Technical Manual 183
Parameter Range Standard
Shock Path (Valid
for all shocks
including the pain-
free shock
impedance)
RV -> SVC + ICD
(Housing)
RV -> ICD (500/540 only)
RV -> SVC (500/540
only)
RV -> SVC +
ICD
Pulse Amplitudes and Pulse Widths
Pulse amplitude 0.2...(0.1)...6.2, 7.5 V 2.8 V
Pulse amplitude
(Lumax 700/740 &
600/640 models
only)
0.5…(0.25)…4.0…(0.5)…
..6.0, 7.5 V
2.5 V
Pulse width 0.4, 0.5, 0.7, 1.0, 1.2, 1.5
ms
0.4 ms
Pulse width
(Lumax 700/740 &
600/640 models
only)
0.4, 0.5…(0.25)…1.5 ms 0.4 ms
Automatic Threshold Measurement (ATM)
For Lumax 500/540 Models Only
RV ON; OFF ON
1
LV (HF-T Only) ON; OFF ON
Active Threshold Tracking (ATT)
For Lumax 700/740 & 600/640 Models Only
RV ON; OFF ON
2
LV (HF-T Only) ON; OFF ON
Thoracic Impedance Measurement (TI)
For Lumax 700/740 & 600/640 Models Only
TI ON; OFF ON
Mode
Modes DDD, DDDR, DDI, DDIR,
VDD, VDDR, VDI, VDIR,
VVI, VVIR, AAI, AAIR,
OFF
DDD
1
The Standard Program for the ATM feature is ON only for the
Lumax 500/540 VR-T
2
The Standard Program for the ATT feature is ON only for the
Lumax 700/740 & 600/640 VR-T