User Manual

Table Of Contents
Lumax Technical Manual 181
Parameter Range Standard
Hold of upper
threshold
100...(20)...600 ms 360 ms
Hold of upper
threshold after Vs
(Lumax 700/740 &
600/640 only)
110, 150...(50)...500 ms
110 ms (VFS)
350 ms
Hold of upper
threshold after Vp
(Lumax 700/740 &
600/640 only)
110, 150...(50)...500 ms
110 ms (VFS)
400 ms
Initial Noise Interval
(Lumax 700/740 &
600/640 only)
110 …(10) … 200 ms 110 ms
High pass
(Lumax 700/740 &
600/640 only)
24, 32 Hz
32 Hz (TWS, VFS)
24 Hz
Low pass
(Lumax 700/740 &
600/640 only)
100 Hz 100 Hz
Left-ventricular Sensing Parameters
Sensing LV STD - standard,
OFF - inactive
STD
Min. threshold 0.5...(0.1)...2.5 mV 1.6 mV
Min. threshold
(Lumax 700/740 &
600/640 only)
0.5…(0.1)…2.5….(0.5)….
5.0 mV
1.6 mV
Blanking after atrial
pacing
= Safety window 100 ms
(N/A
VR-T DX)
Upper threshold 50; 75; 87.5% 50 %
Upper threshold
(Lumax 700/740 &
600/640 only)
50; 75 % 50 %
Hold of upper
threshold
100...(20)...600 ms 360 ms
Hold of upper
threshold after Vs
(Lumax 700/740 &
600/640 only)
110, 150...(50)...500 ms 350 ms