Lumax Technical Manual 179
6. Technical Specifications
The following are the technical specifications for the
Lumax ICDs/CRT-Ds. The ranges are presented in the format:
x…(y)…z
where x = the lowest value, y = the increment, and z = the largest
value.
Mechanical Properties
Lumax
300/500
600/700
Lumax
340/540
640/740
Lumax
500
600/700
HF-T
Lumax
340/540
640/740
HF-T
Dimensions 66 mm x
55 mm x
12 mm
66 mm x
55 mm x
13 mm
66 mm x
59 mm x
12 mm
66 mm x
59 mm x
13 mm
Volume 34.6 cm
3
37.2 cm
3
37.1 cm
3
39.8 cm
3
Mass 81 g 92 g 83 g 94 g
Housing Titanium
Header Epoxy resin
Seal Plug Silicone
VR-T
Lead Ports
1 x 3.2 mm IS-1 Bipolar
2 x 3.2 mm DF-1
DR-T
Lead Ports
2 x 3.2 mm IS-1 Bipolar
2 x 3.2 mm DF-1
3 x 3.2 mm IS-1 Bipolar
2 x 3.2 mm DF-1
Parameter Range Standard
Bradycardia
Atrial Sensing Parameters
Sensing STD - standard,
OFF - inactive
STD
Minimum threshold 0.2...(0.1)...2.0 mV 0.4 mV
Far-field protection
after Vp
50...(25)...225 ms 75 ms
Far-field protection
after Vs
Off, 25...(25)...225 ms 75 ms
Upper threshold 50; 75; 87.5 % 50 %