174 Lumax Technical Manual
5.2.3 Lumax 600/640 & 700/740 Devices
Table 41 provides longevity estimates for the single chamber
Lumax 600/640 & 700/740 ICD variants (Lumax 600/700 VR-T
and Lumax 640/740 VR-T). The table provides several different
support scenarios. It is assumed that the shocks are equally
spaced throughout the life of the ICD.
Table 46 Lumax VR-T Longevity Estimates
Pacing Support
Device Shocks per
year
0% 15% 50% 100%
4 9.91 9.60 8.95 8.17
8 8.32 8.10 7.64 7.06
600 VR-T
700 VR-T
12 7.17 7.01 6.66 6.21
4 10.21 9.92 9.31 8.55
8 8.20 8.01 7.60 7.09
640 VR-T
740 VR-T
12 6.84 6.71 6.42 6.05
Table 42 provides longevity estimates for the dual-chamber
Lumax 600/640 and 700/740 ICD variants (Lumax 600/700 DR-T,
and Lumax 640/740 DR-T). The table provides several different
support scenarios. It is assumed that the shocks are equally
spaced throughout the life of the ICD.
Table 47 Lumax DR-T Longevity Estimates
Pacing Support
Device Shocks per
year
0% 15% 50% 100%
4 8.95 8.42 7.40 6.31
8 7.62 7.24 6.47 5.61
600 DR-T
700 DR-T
12 6.64 6.34 5.74 5.06
4 9.31 8.80 7.80 6.71
8 7.59 7.25 6.55 5.77
640 DR-T
740 DR-T
12 6.40 6.16 5.65 5.05
Table 43 provides longevity estimates for the
Lumax 640/740 VR-T DX ICDs. The table provides several
different support scenarios. It is assumed that the shocks are
equally spaced throughout the life of the ICD.