170 Lumax Technical Manual
Table 38 Lumax 340 DR-T Longevity Estimates
Pacing Support
Shocks per year
(max energy)
0% 15% 50% 100%
4 5.7 5.4 4.8 4.1
5 5.4 5.1 4.6 3.9
6 5.2 4.9 4.4 3.8
7 5.0 4.7 4.2 3.7
8 4.8 4.5 4.1 3.6
9 4.6 4.4 3.9 3.4
10 4.4 4.2 3.8 3.3
11 4.2 4.0 3.7 3.2
12 4.1 3.9 3.6 3.2
Table 39 and Table 40 provide longevity estimates for the
Lumax 300 & 340 HF-T CRT-D with Greatbatch batteries. The
tables are limited to 100% ventricular pacing because that is the
goal for each CRT-D device. It is assumed that the shocks are
equally spaced throughout the life of the ICD.
Table 39 Lumax 300 HF-T Longevity Estimates
Shocks per year 100% Pacing Support
(in years)
4 5.4
5 5.3
6 5.2
7 5.1
8 4.9
9 4.8
10 4.7
11 4.6
12 4.5