Lumax Technical Manual 165
Prior to surgically closing the pocket, the telemetry contact should
be evaluated to help ensure chronic programmer communication.
Close the device pocket using standard surgical technique. As
the final step at device implant and each patient follow-up, the
permanent program should be retransmitted to the ICD/CRT-D.
Typically, each device that you receive is in the “Shipment Mode”.
This mode includes factory settings that control the charge
current of automatic capacitor reformations to avoid the possibility
of temporary low battery readings. Shipment mode is
automatically deactivated when electrophysiological tests (e.g.,
Impedance Measurement) are initiated by the programmer. The
following can be used to verify status of the shipment mode:
The shipment mode is ON if the device displays
"Shipment Mode Active" in the event list
The Shipment Mode is OFF if the device does not display
"Shipment Mode Active" in the event list
Complete the Medical Device Registration Form provided with the
ICD/CRT-D and return it to BIOTRONIK.