User Manual
Table Of Contents
- 1. General
- 1.1 System Description
- 1.2 Indications and Usage
- 1.3 Contraindications
- 1.4 Warnings and Precautions
- 1.4.1 Sterilization, Storage, and Handling
- 1.4.2 Device Implantation and Programming
- 1.4.3 Lead Evaluation and Connection
- 1.4.4 Follow-up Testing
- 1.4.5 Pulse Generator Explant and Disposal
- 1.4.6 Hospital and Medical Hazards
- 1.4.7 Home and Occupational Hazards
- 1.4.8 Cellular Phones
- 1.4.9 Electronic Article Surveillance (EAS)
- 1.4.10 Home Appliances
- 1.4.11 Home Monitoring®
- 1.5 Potential/Observed Effects of the Device on Health
- 1.6 Clinical Studies
- 1.6.1 Kronos LVT Study
- 1.6.2 Tupos LV/ATx Study
- 1.6.2.1 Study Overview
- 1.6.2.2 Methods
- 1.6.2.3 Summary of Clinical Results
- 1.6.2.4 Primary Endpoint 1: Six Minute Walk Test & QOL (Effectiveness)
- 1.6.2.5 Effectiveness Endpoint Analysis and Conclusions
- 1.6.2.6 Primary Endpoint 2: Complication-Free Rate (Safety)
- 1.6.2.7 Primary Safety Enpoint Analysis and Conclusions
- 1.6.2.8 Post-hoc Safety Analysis
- 1.6.2.9 Post hoc Safety Analysis Conclusion
- 1.6.2.10 Secondary Endpoint Results
- 1.6.2.11 Multi-site Poolability and Gender Analysis
- 1.6.2.12 Conclusions
- 1.6.3 Lumax HFT VV Clinical Study
- 1.6.4 TRUST Clinical Study
- 1.6.5 Deikos A+
- 1.7 Patient Selection and Treatment
- 1.8 Patient Counseling Information
- 1.9 Evaluating Prospective CRTD/ICD Patients
- 2. Device Features
- 2.1 SafeSync Telemetry
- 2.2 Cardiac Resynchronization Therapy (CRT)
- 2.3 Sensing (Automatic Sensitivity Control)
- 2.4 Automatic Threshold Measurement (ATM)
- 2.5 Ventricular Tachyarrhythmia Detection
- 2.6 Tachyarrhythmia Redetection
- 2.7 Tachyarrhythmia Termination
- 2.8 Tachyarrhythmia Therapy
- 2.9 Bradycardia Therapy
- 2.9.1 Bradycardia Pacing Modes
- 2.9.2 Basic Rate
- 2.9.3 Night Rate
- 2.9.4 Rate Hysteresis
- 2.9.5 Dynamic AV Delay
- 2.9.6 IOPT
- 2.9.7 Upper Tracking Rate
- 2.9.8 Mode Switching
- 2.9.9 PMT Management
- 2.9.10 VES Discrimination after Atrial Sensed Events
- 2.9.11 Rate-Adaptive Pacing
- 2.9.12 Pulse Amplitude
- 2.9.13 Pulse Width
- 2.9.14 Post Ventricular Atrial Refractory Period
- 2.9.15 PVARP after VES
- 2.9.16 Auto PVARP
- 2.9.17 Noise Response
- 2.9.18 Post Shock Pacing
- 2.10 EP Test Functions
- 2.11 Special Features
- 2.10.2.3 Transmitting Data
- 2.11.3.3 Types of Report Transmissions
- 2.11.3.4 Description of Transmitted Data
- 2.11.3.5 IEGM Online HDs
- 2.11.3.6 Scheduling Remote Follow-up
- 2.11.4 Real-time IEGM Transmission
- 2.11.5 Capacitor Reforming
- 2.11.6 Patient and Implant Data
- 2.11.7 System Status
- 2.11.8 HF Monitor Statistics
- 2.11.9 Holter Memory
- 2.11.10 Timing Statistics
- 2.11.11 Atrial Arrhythmias
- 2.11.12 Ventricular Arrhythmias
- 2.11.13 Sensor
- 2.11.14 Sensing
- 2.11.15 Impedances
- 2.11.16 Automatic Threshold
- 2.11.17 Asynchronous Pacing Modes
- 2.11.18 Far-Field IEGM for Threshold Testing (Leadless ECG)
- 2.11.19 Advanced AT/AF Diagnostics (Lumax 700/740 only)
- 2.11.20 Atrial NIPS (Lumax 700/740 & 600/640 only)
- 3. Sterilization and Storage
- 4. Implant Procedure
- 5. Follow-up Procedures
- 6. Technical Specifications
Lumax Technical Manual 161
1. Confirm that the setscrews are not protruding into the
connector receptacles. To retract a setscrew, insert the
enclosed torque wrench through the perforation in the
self-sealing plug at an angle perpendicular to the lead
connector until it is firmly placed in the setscrew. Rotate
the wrench counterclockwise until the receptacle is clear
of obstruction.
2. Insert the lead connector into the connector port of the
ICD/CRT-D without bending the lead until the connector
pin becomes visible behind the setscrew. Hold the
connector in this position. If necessary, apply silicone oil
only to the o-rings on the connector (not the connector
pin).
3. Insert the enclosed torque wrench through the perforation
in the self-sealing plug at an angle perpendicular to the
lead connector until it is firmly placed in the setscrew.
4. Securely tighten the setscrew of the connector clockwise
with the torque wrench until torque transmission is limited
by the wrench.
5. Carefully retract the torque wrench. The perforation will
self-seal.
4.6 Blind Plug Connection
The Lumax DR ICD and HF CRT-D are shipped with a blind plug
(pre inserted) in an unused header port. Refer to the following
steps when connecting blind plugs to the device.
1. Confirm that the setscrews are not protruding into the
connector receptacles. To retract a setscrew, insert the
enclosed torque wrench through the perforation in the
self-sealing plug at an angle perpendicular to the lead
connector until it is firmly placed in the setscrew. Rotate
the wrench counterclockwise until the receptacle is clear
of obstruction.
2. Insert the blind plug into the connector port of the
ICD/CRT-D until the connector pin becomes visible
behind the setscrew.
3. Insert the enclosed torque wrench through the perforation
in the self-sealing plug at an angle perpendicular to the
connector until it is firmly placed in the setscrew.
4. Securely tighten the setscrew of the connector clockwise
with the torque wrench until torque transmission is limited
by the wrench.