User Manual

Table Of Contents
Lumax Technical Manual 7
Insertion and withdrawal forces as specified by
ISO 5841-3 (IS-1) and ISO 11318:1993 (E) DF-1
The following leads were evaluated in a retrospective study with
BIOTRONIK’s ICDs/CRT-Ds:
Medtronic Sprint
TM
Lead 6932
Medtronic Sprint Lead 6943
Medtronic Sprint Quattro
TM
Lead 6944
Medtronic Transvene
TM
RV Lead 6936
St. Jude (Ventritex) TVL
TM
- ADX Lead 1559
St. Jude SPL® SP02 Lead
Guidant ENDOTAK® DSP Lead
Guidant ENDOTAK Endurance EZ Lead, ENDOTAK
Reliance Lead
Guidant (Intermedics) Lead 497-24.
The following leads were bench tested for compatibility with
BIOTRONIK’s ICDs/CRT-Ds:
Guidant ENDOTAK Endurance Lead “CPI 0125”
Guidant ENDOTAK Reliance Lead 0148
Medtronic Sprint Lead 6932
Medtronic Sprint Lead 6942
Medtronic Sprint Lead 6943
Medtronic Sprint Lead6945
Medtronic Sprint Quattro Lead 6944
St. Jude Riata® Lead 1571/65
St. Jude SPL SPO1 Lead.
Resuscitation Availability - Do not perform induction testing
unless an alternate source of patient defibrillation such as an
external defibrillator is readily available. In order to implant the
ICD/CRT-D system, it is necessary to induce and convert the
patient’s ventricular tachyarrhythmias.
Unwanted Shocks – Always program ICD Therapy to OFF prior
to handling the device to prevent the delivery of serious shocks to
the patient or the person handling the device during the implant
procedure.
Rate-Adaptive Pacing – Use rate-adaptive pacing with care in
patients unable to tolerate increased pacing rates.