158 Lumax Technical Manual
4.5 Lead to Device Connection
The Lumax ICD/CRT-Ds have been designed and are
recommended for use with a defibrillation lead systems having
one IS-1 connector for ventricular sensing and pacing and up to
two DF-1 connectors for delivery of shock therapy. A separate
bipolar atrial lead with IS-1 connector is required for atrial sensing
and pacing functions and the CS lead for biventricular pacing
(LV). Figure 15 depicts the configuration
of the header ports on
the Lumax 300/340 ICD/CRT-Ds, where HV1/RV and HV2/SVC
are for DF-1 connectors, and A P/S and V P/S are for IS-1
connectors. Figure 16 depicts the configuration of the header
ports on the Lumax 500/540,
600/640 and 700/740 ICD/CRT-D
devices.
VR(-T) DR(-T)/VR-T DX HF(-T)
Figure 15. Lumax 300/340 ICD and CRT-D Header Ports