User Manual

Table Of Contents
Lumax Technical Manual 157
CAUTION
Device Packaging - Do not use the device if the device’s
packaging is wet, punctured, opened or damaged because
the integrity of the sterile packaging may be compromised.
Return the device to BIOTRONIK.
4.4 Pocket Preparation
Using standard surgical technique, create a pocket for the device
either in the patient’s pectoral or abdominal region dependent on
patient anatomy. The device may be implanted either below the
subcutaneous tissue or in the muscle tissue. The ICD/CRT-D
should be implanted with the etched side facing up. The leads
should be tunneled or surgically brought into the device pocket. If
lead tunneling is performed, re-evaluation of the baseline lead
signals, after tunneling is recommended.
CAUTION
The ICD/CRT-D system should have detection and therapy
disabled prior to performing medical procedures. In addition,
the ICD/CRT-D should be checked after the procedures to
assure proper programming:
Electrocautery - Electrosurgical cautery could induce
ventricular arrhythmias and/or fibrillation, or may cause
device malfunction or damage. If use of electrocautery is
necessary, the current path and ground plate should be kept
as far away from the pulse generator and leads as possible
(at least 6 inches (15 cm)).