154 Lumax Technical Manual
WARNING
Left Ventricular Lead Systems – BIOTRONIK CRT-Ds
maybe implanted with any legally marketed, compatible LV
lead. Compatibility is defined as:
IS-1 pacing connector
Active or passive fixation technology
Insertion and withdrawal forces as specified by
ISO 5841-3 (IS-1)
The following LV leads were evaluated in the
OPTION CRT/ATx study with BIOTRONIK’s CRT-Ds:
Guidant-EASYTRAK® IS-1 Lead
Guidant-EASYTRAK LV-1 Lead
Guidant-EASYTRAK 2 Lead
Guidant-EASYTRAK 3 Lead
Medtronic-Attain® OTW Lead
St. Jude-Aescula
TM
Lead
St. Jude-QuickSite® Lead
Biomec-Myopore
TM
Epicardial Lead
Medtronic-Epicardial 5071 Lead
Medtronic-CapSure® EPI Lead
BIOTRONIK-ELC 54-UP Lead
The following LV leads were bench tested for compatibility
with BIOTRONIK’s CRT-Ds:
Guidant EASYTRAK 4512 (unipolar) Lead
Guidant EASYTRAK 4513 (bipolar) Lead
Guidant EASYTRAK 3 4525 (bipolar) Lead
Medtronic Attain OTW 4193 (unipolar) Lead
Medtronic Attain OTW 4194 (bipolar) Lead
Medtronic Attain LV 2187 (unipolar) Lead
St. Jude Medical QuickSite 1056K (unipolar) Lead
ELA SITUS® OTW (unipolar) Lead
BIOTRONIK Corox+ LV-H 75-BP #346542
BIOTRONIK Corox OTW 75-UP Steroid #346542
(unipolar)