2.11.13 Sensor
This report contains individual rate bins for the sensor indicated
rate percentages. The activity report provides information that can
assist the physician in optimizing pacing and/or sensor
parameters.
2.11.14 Sensing
The activity report provides trend information on atrial and right
ventricular sensing measurements.
2.11.15 Impedances
The activity report provides trend information on atrial and right
ventricular pacing impedance as well as shock impedance
measurements.
2.11.16 Automatic Threshold
The activity report provides trend information on right and left
ventricular pacing thresholds, if active.
2.11.17 Asynchronous Pacing Modes
The Lumax 700/740 and 600/640 models offer the following
asynchronous pacing modes for use during medical procedures:
V00 – asynchronous pacing in the ventricle
D00 – asynchronous pacing in the atrium and ventricle
with a fixed AV delay for conduction between chambers
Tachyarrhythmia detection is deactivated when using these
asynchronous modes in the Lumax 700/740 and 600/640
devices.
The asynchronous modes are intended for use during medical
procedures, such as cautery. In patients with inadequate intrinsic
rhythm, the pacemaker should be reprogrammed to an
asynchronous mode during the procedure in order to prevent
inhibition by electromagnetic interference. Thus, the
asynchronous modes V00 and D00 are intended to prevent
possible inhibition by electromagnetic interference during invasive
intervention (such as during electrocauterization).