The MHRR value is measured during a programmed period,
defined by a Rest Period Start Time and a Rest Period Duration.
The resting period shall be adjustable via the programmer. The
MHRR has a value range of 0 to 180 bpm. The daily value
(MHRR) is stored for a period of 240 days. After 240 days, new
daily values replace the oldest daily values. The programmer
presents the daily bpm-value in a trend graph for the last 240
days.
2.11.8.4 Heart Rate Variability
Heart
Rate Variability, which is the standard deviation of the 5-
minute mean normal to normal interval over the recorded time is
available in the statistics of the Lumax devices. This is based on
the atrial rate (P-P).
2.11.9 Holter Memory
Various device information is available within the Holter memory.
The Holter memory can be configured a number of different ways
depending on the physician’s preference.
2.11.9.1 Episode List
The ICD/CRT-D stores a variety
of useful diagnostic data about
tachyarrhythmia episodes, which may be used to optimize
tachyarrhythmia detection and therapy parameters. This
diagnostic data includes detection counters; therapy counters,
last delivered ATP and shock therapy, shock data memory,
therapy history, and stored intracardiac electrograms.
Episode Details
Detailed information about each individual episode presented as
a table of events ordered from most recently delivered to the first
delivered. Each IEGM segment can be viewed from the episode
detail sub-menu by selecting the IEGM button (icon). From this
screen, an IEGM can be expanded and scrolled to assist in a
more accurate IEGM interpretation by enabling a closer
examination of specific segments.