140 Lumax Technical Manual
CAUTION
Capacitor Reformation - Infrequent charging of the high
voltage capacitors may extend the charge times of the
ICD/CRT-D. The capacitors are automatically reformed.
2.11.6 Patient and Implant Data
The Patient and Implant data screens allow input of data
regarding the patient name, demographics, implanting physician,
date, devices implanted, location of the implant, and various
conditions related to the patient. This information is transmitted to
the ICD/CRT-D and resides in the device memory for later recall if
needed.
2.11.7 System Status
Various device parameters can be monitored through the Status
section of the programmer screen. Displayed data includes
ICD/CRT-D information, charge circuit parameters, capacitor
reform information, battery status and voltage, and lead
information. The system status screen presents a large variety of
information about the Lumax ICDs/CRT-Ds including:
Serial number (always displayed after interrogation)
Software Release
ICD status
PID Number
Battery status
Battery voltage
Last charge event and Last event with a maximum
energy charge:
Energy
Charge time
Date
Time
Total number of charges
Last Home Monitoring message
Type of message
Time and date