User Manual

Table Of Contents
138 Lumax Technical Manual
2.11.3.6 Scheduling Remote Follow-up
The Lumax 700/740 and 600/640 ICDs/CRT-Ds provide the
ability to automatically schedule remote follow-ups using the
programmer. This feature has the following two options:
1. Fixed Follow-up Intervals (using 30-180 day cycles) from
the date of the first scheduled transmission:
2. Selection of 5 sequential dates* with a minimum time lag
of 20 days between any two selected dates.
*If the 5 dates have been expired the standard period for periodic
IEGM (30 days) will be enabled.
2.11.4 Real-time IEGM Transmission
The pulse generators provide real time transmission of the
unfiltered intracardiac electrogram (IEGM) to the programmer.
IEGMs from the atrium and ventricles can be simultaneously
recorded with a bandwidth of 0.5 to 200 Hz. Depending on the
device, the following channels are simultaneously recorded:
During single chamber (VR/VR-T) operation, far field, and
RV electrograms are available.
During dual chamber (DR/DR-T) operation, far field, RA,
and RV electrograms are available.
During triple chamber (HF/HF-T) operation, RA, RV, and
LV electrograms are available.