132 Lumax Technical Manual
Periodic Report with IEGM
The Lumax ICDs and CRT-Ds can be programmed to transmit
IEGMs with the daily Home Monitoring report on a periodic basis
with the interval selected; OFF, 2, 3, 4, or 6 months.
Event Report
When certain cardiac and technical events occur, a report is
automatically generated. This information is described as an
“event report.”
The implant supports the following automatic event triggers:
Termination of VT/VF Episode (not Termination of a
Monitoring episode, SVT episode or Detection while
Magnet applied (DTM) episode)
Ongoing Atrial Monitoring Episode lasting longer than the
programmed time (0.5, 6, 12 or 18 hours)
Impedance out-of-range for A, RV, LV, shock lead
(painless)
Special Device Status
First ineffective 30 J shock detected
Initial detection SVT
Percent of ventricular sensing below set limit
Device status – ERI/EOS/ROM-Mode
Programmer Triggered Report
With the device status screen, it is possible to test the
ICD/CRT-D’s Home Monitoring capabilities during device
implantation or follow-up.
NOTE:
Battery voltage and pace/sense lead impedance are
measured before the first transmission of the day. Therefore,
the first transmission may occur 2 minutes after the
programmed transmission time.
2.11.3.4 Description of Transmitted Data
The following data are transmitted for the Cardio Report by the
Home Monitoring system, when activated. In addition to the
medical data, the serial number of the implant is also transmitted.