2.11.3.3 Types of Report Transmissions
When
the Home Monitoring function is activated, the transmission
of information from the implant can be triggered as follows:
Daily report – the time period (daily) initiates the
transmission (IEGM included if not yet sent)
Event report – the ICD/CRT-D detects certain events,
which initiate a transmission (IEGM included if not yet
sent)
Periodic IEGM report – certain event reports can be
programmed to have an IEGM included each time that
they are transmitted. The IEGM includes comprehensive
event details with up to 10 seconds of IEGM (8 seconds
pre-detection and 2 seconds pre-termination).
Programmer triggered report – a test message
transmitted upon request of the physician in the clinic.
Daily Report
The time of daily Home Monitoring Report transmission is
programmable. For periodic messages, the time can be set
anywhere between 0:00 and 23:59 hours. It is recommended to
select a time between 0:00 and 4:00.