User Manual

Table Of Contents
Lumax Technical Manual 125
2.10.9 Manual ATP
The ICD/CRT-D can deliver a manual ATP on demand through a
programmer command in the EP test menu. To deliver an ATP
sequence, place the wand over the device and select the Start
ATP button. A confirmation menu will appear and the
programmed pacing sequence command will be delivered upon
selecting the OK button in this screen. Programming of the
manual ATP is similar to the programming available for automatic
ATP therapy as described in Section
2.8.2.
2.10.10 Emergency Shock
The ICD/CRT-D can deliver an emergency maximum energy
shock on demand through a programmer command in the EP test
menu. For emergency shocks, the energy (maximum 30 or 40 J),
polarity (Standard) and waveform (Biphasic) are pre-defined. To
deliver a shock, place the wand over the device and select the
Emergency Shock button. A confirmation menu will appear and
the shock will be delivered.
2.11 Special Features
The Lumax includes several special features to improve ease of
use and provide additional information to the user.
2.11.1 ICD Therapy Status
Interrogate the device to observe the ICD Therapy window (upper
right hand corner) of the main programming screen which
indicates the ICD Therapy status (either ON or OFF). The overall
ICD therapy can be controlled by selecting ON (enabled) or OFF
(disabled). If the device has not been interrogated, the
programmer displays “Unknown” highlighted in red in the ICD
therapy window.