User Manual
Table Of Contents
- 1. General
- 1.1 System Description
- 1.2 Indications and Usage
- 1.3 Contraindications
- 1.4 Warnings and Precautions
- 1.4.1 Sterilization, Storage, and Handling
- 1.4.2 Device Implantation and Programming
- 1.4.3 Lead Evaluation and Connection
- 1.4.4 Follow-up Testing
- 1.4.5 Pulse Generator Explant and Disposal
- 1.4.6 Hospital and Medical Hazards
- 1.4.7 Home and Occupational Hazards
- 1.4.8 Cellular Phones
- 1.4.9 Electronic Article Surveillance (EAS)
- 1.4.10 Home Appliances
- 1.4.11 Home Monitoring®
- 1.5 Potential/Observed Effects of the Device on Health
- 1.6 Clinical Studies
- 1.6.1 Kronos LVT Study
- 1.6.2 Tupos LV/ATx Study
- 1.6.2.1 Study Overview
- 1.6.2.2 Methods
- 1.6.2.3 Summary of Clinical Results
- 1.6.2.4 Primary Endpoint 1: Six Minute Walk Test & QOL (Effectiveness)
- 1.6.2.5 Effectiveness Endpoint Analysis and Conclusions
- 1.6.2.6 Primary Endpoint 2: Complication-Free Rate (Safety)
- 1.6.2.7 Primary Safety Enpoint Analysis and Conclusions
- 1.6.2.8 Post-hoc Safety Analysis
- 1.6.2.9 Post hoc Safety Analysis Conclusion
- 1.6.2.10 Secondary Endpoint Results
- 1.6.2.11 Multi-site Poolability and Gender Analysis
- 1.6.2.12 Conclusions
- 1.6.3 Lumax HFT VV Clinical Study
- 1.6.4 TRUST Clinical Study
- 1.6.5 Deikos A+
- 1.7 Patient Selection and Treatment
- 1.8 Patient Counseling Information
- 1.9 Evaluating Prospective CRTD/ICD Patients
- 2. Device Features
- 2.1 SafeSync Telemetry
- 2.2 Cardiac Resynchronization Therapy (CRT)
- 2.3 Sensing (Automatic Sensitivity Control)
- 2.4 Automatic Threshold Measurement (ATM)
- 2.5 Ventricular Tachyarrhythmia Detection
- 2.6 Tachyarrhythmia Redetection
- 2.7 Tachyarrhythmia Termination
- 2.8 Tachyarrhythmia Therapy
- 2.9 Bradycardia Therapy
- 2.9.1 Bradycardia Pacing Modes
- 2.9.2 Basic Rate
- 2.9.3 Night Rate
- 2.9.4 Rate Hysteresis
- 2.9.5 Dynamic AV Delay
- 2.9.6 IOPT
- 2.9.7 Upper Tracking Rate
- 2.9.8 Mode Switching
- 2.9.9 PMT Management
- 2.9.10 VES Discrimination after Atrial Sensed Events
- 2.9.11 Rate-Adaptive Pacing
- 2.9.12 Pulse Amplitude
- 2.9.13 Pulse Width
- 2.9.14 Post Ventricular Atrial Refractory Period
- 2.9.15 PVARP after VES
- 2.9.16 Auto PVARP
- 2.9.17 Noise Response
- 2.9.18 Post Shock Pacing
- 2.10 EP Test Functions
- 2.11 Special Features
- 2.10.2.3 Transmitting Data
- 2.11.3.3 Types of Report Transmissions
- 2.11.3.4 Description of Transmitted Data
- 2.11.3.5 IEGM Online HDs
- 2.11.3.6 Scheduling Remote Follow-up
- 2.11.4 Real-time IEGM Transmission
- 2.11.5 Capacitor Reforming
- 2.11.6 Patient and Implant Data
- 2.11.7 System Status
- 2.11.8 HF Monitor Statistics
- 2.11.9 Holter Memory
- 2.11.10 Timing Statistics
- 2.11.11 Atrial Arrhythmias
- 2.11.12 Ventricular Arrhythmias
- 2.11.13 Sensor
- 2.11.14 Sensing
- 2.11.15 Impedances
- 2.11.16 Automatic Threshold
- 2.11.17 Asynchronous Pacing Modes
- 2.11.18 Far-Field IEGM for Threshold Testing (Leadless ECG)
- 2.11.19 Advanced AT/AF Diagnostics (Lumax 700/740 only)
- 2.11.20 Atrial NIPS (Lumax 700/740 & 600/640 only)
- 3. Sterilization and Storage
- 4. Implant Procedure
- 5. Follow-up Procedures
- 6. Technical Specifications
Lumax Technical Manual 93
S1-S2 Interval - The S1-S2 programmable coupling interval
occurs between the Burst sequence and the extra stimuli (PES). It
defines the interval between the first stimulus (S1) and the extra
stimuli (S2).
Scan Decrement - The Scan decrement continuously reduces
the starting pulse intervals of each Burst or Scan.
Minimum ATP Interval – For the Lumax 300/340 & 500/540
models, the programmed minimum interval prevents ATPs from
being given with stimulation values less than the minimum
interval. When the ATP interval reaches the value of the minimum
interval with the S1 decrement or scan decrement, it then
assumes this value and remains constant. For the Lumax
600/640 & 700/740 models, the minimum ATP interval is 200 ms
fixed.
2.8.2.1 ATP Help
ATP help is a useful tool to assist
the physician in choosing and
confirming appropriate ATP programming. The “ATP Help” button
is displayed in each ATP therapy option. When the ATP help
button is pressed, a histogram of the chosen therapy scheme is
shown. The histogram displays the intervals for the programmed
ATP scheme. When rate adaptive intervals are programmed, the
displayed intervals are based on the programmed R-R average.
2.8.2.2 ATP Optimization
In
order to optimize future therapies, the ICD will store the
parameter configuration of the last successful ATP attempt in
each the VT1 and VT2 classes. The last successful stored ATP
attempt is then used as the starting point for the next detected
episode of the same arrhythmia class. If the stored parameter
configuration is not successful, it is deleted from the ATP
optimization memory of the respective arrhythmia class and
subsequent therapy sequences will begin with ATP1 for the next
detected episode.
ATP Optimization is programmable ON or OFF for all ATP
therapies and VT zones with one parameter.