User's Manual

61
Pacing and Sensing Functions
CLS Standby Mode
CLS Standby Mode is where, after auto-initialization, Closed Loop
Stimulation is fully installed but deactivated. When the implant is
interrogated for the first time, Closed Loop Stimulation can be
activated. The CLS Standby Mode entails the following:
Closed Loop Stimulation has been installed, but is deactivated.
Until CLS is activated, the implant will pace at the basic rate.
During the first implant interrogation, the user can activate
Closed Loop Stimulation.
CLS is activated
Activating CLS after the first interrogation means the following
parameters have been set:
DDD-CLS mode for dual-chamber implants or VVI-CLS mode for
single-chamber implants.
AV delays and AV hystereses are automatically optimized for
CLS.
CLS is not activated
When Closed Loop Stimulation has not been activated after the first
follow-up, the following parameters are automatically set:
DDD mode for dual-chamber implants or VVI mode for single-
chamber implants.
The AV delays from the factory program are activated. AV
hystereses are turned off.
Programmable Parameters
In addition to activating and deactivating the entire function, the sub-
functions of lead detection can be activated individually.
Note: Patient data can always be configured regardless
of the auto-initialization function.