User's Manual

60
Pacing and Sensing Functions
Figure 11: Implant functions that are activated by auto-initialization
During auto-initialization, the implant activates the following functions:
Statistics
ATM (the threshold recording feature of the ACC function)
Mode switching
PMT management
Closed Loop Stimulation standby mode, meaning that the
function has been completely initialized and deactivated
Note: If the implant parameters have been changed in
the factory program prior to implantation, the auto-
initialization function can no longer be executed. In
this case, only the lead detection function can be
run. Exception: Patient data can always be
configured regardless of the auto-initialization
function.
Note: The lead detection and auto-initialization functions
can be run in the ventricle only with single-
chamber pacemakers.
select lead polarity
implantation confirmation 30
min.
standby
activating functions
statistics
PMT
protection
threshold
monitoring