User's Manual
58
Pacing and Sensing Functions
Lead Detection and Auto-Initialization
Lead Detection
Purpose
The lead detection function allows the implant to recognize the
connected leads as early as during implantation. This is also the basis
for being able to activate the auto-initialization function.
When the connected leads have been successfully detected, the pacing
and sensing polarities are automatically set. This depends on the type
of leads connected (be they unipolar or bipolar). The pacemaker uses
the lead impedance as a basis for the automatic polarity setting.
The pacemaker goes through the following phases:
— An initial lead detection
— Lead polarity is recognized
— Confirmation
The Initial Lead Detection
To detect a lead, the implant (depending on the type) provides unipolar
pacing both in the atrial and ventricular channel and measures the
impedance of each stimulus. If intrinsic events are detected, they
trigger a pulse in the same chamber in which the event was detected.
This mimics the pacing response of an implant in the DDT mode. If the
measured impedances lie within 200-3000 Ohm, the lead is considered
detected.
Recognizing Lead Polarity
After successful detection of the lead, the implant switches to bipolar
pacing. The impedance is also measured during pacing. If it lies
between 200-3000 Ohm, a bipolar lead is considered confirmed.