User's Manual

38
Timing Functions
Basic Rate:
Increased value, for example 90 ppm
Rate Hysteresis:
Such that the hysteresis rate at rest is always lower than the intrinsic
rhythm (e.g., -50)
Scan Rate Hysteresis:
Enabled, with the number of cycles set according to the patient's
condition
Repetitive rate
hysteresis
Enabled, with a low number of cycles
This programming will inhibit the pacemaker until bradycardia
episodes occur. If the rate drops due to an event, the pacemaker will
pace at the hysteresis rate for the set number of repetition cycles (the
confirmation period). The pacemaker will switch to the higher
intervention rate to prevent possible syncope only if a spontaneous
rhythm does not occur during the confirmation period, which should be
set as short as possible. The pacemaker will scan for a spontaneous
rhythm every 180 cycles (scan rate hysteresis) to avoid long pacing
phases. If the attack has been terminated by that time, the pacemaker
will be inhibited; otherwise, it will repeat the scan every 180 cycles.
Note: These patients should only be treated with a
DDD(R) system to exploit the contribution of the
atrium to ventricular filling and to overall
hemodynamics as much as possible during such
attacks.
Night Program
When the night program is activated, the pacemaker reduces its
activity during the night. This makes it possible to adapt the pacing
rate to the patient's reduced metabolic needs during this time.
Furthermore, VVI and VOO pacing may prevent the possible worsening
of hemodynamics.