User's Manual

161
Technical Data
Cylos VR
AV Delay
Basic rate
70 ppm
70-90 ppm
91-110 ppm
111-130 ppm
130 ppm
180 +15/-5 ms
180 +15/-5 ms
160 +15/-5 ms
140 +15/-5 ms
120 +15/-5 ms
100 +15/-5 ms
Atrium
Ventricle
Pulse Amplitude
maximum value
EN 455002-2-1
mean value
3.6 +0.1/-0.7
3.3 +0.1/-0.7
Pulse width 0.42 ± 0.02 ms
Sensitivity
15 ms sin
2
40 ms sin
2
0.2 + 0.05/-0.1 mV
2.5 ± 0.5 mV
EN 455002-2-1
delta pulse
0.24 + 0.05/-0.1 mV 2.5 ± 0.5 mV
Refractory period 425 +10/-20 ms 250 +10/-20 ms
Runaway protection 200 +20/-5 min
-1
Product Line
Model Lead Connection Catalog Number
Cylos DR
uncoated
coated
IS-1
IS-1
349799
349804
Cylos DR-T
uncoated
coated
IS-1
IS-1
349806
349810
Cylos VR
uncoated
coated
IS-1
IS-1
341824
341815