User's Manual

160
Technical Data
Projected Tolerances of Factory Settings
1
Data according to EN 455002-2-1
Cylos DR/DR-T
Basic rate
Interference rate
60 ± 1.5 min
-1
Basic Interval 1000 ± 20 ms
Escape Interval 1000 ± 20 ms
Magnet Rate 90 ± 3 min
-1
(for 10 cycles)
Magnet Interval 664 ± 20 ms (for 10 cycles)
AV Delay
Basic rate
70 ppm
70-90 ppm
91-110 ppm
111-130 ppm
130 ppm
180 +15/-5 ms
180 +15/-5 ms
160 +15/-5 ms
140 +15/-5 ms
120 +15/-5 ms
100 +15/-5 ms
Atrium
Ventricle
Pulse Amplitude
maximum value
EN 455002-2-1
mean value
3.6 +0.1/-0.7
3.3 +0.1/-0.7
3.6 +0.1/-0.7
3.3 +0.1/-0.7
Pulse width 0.42 ± 0.02 ms 0.42 ± 0.02 ms
Sensitivity
15 ms sin
2
40 ms sin
2
1.0 ± 0.5 mV
2.5 ± 0.5 mV
EN 455002-2-1
delta pulse
1.8 ± 0.5 mV 2.5 ± 0.5 mV
Refractory period 425 +10/-20 ms 250 +10/-20 ms
Runaway protection 200 +20/-5 min
-1
200 +20/-5 min
-1
Cylos VR
Basic rate
Interference rate
60 ± 1.5 min
-1
Basic Interval 1000 ± 20 ms
Escape Interval 1000 ± 20 ms
Magnet Rate 90 ± 3 min
-1
(for 10 cycles)
Magnet Interval 664 ± 20 ms (for 10 cycles)
1
37°C, 500