User's Manual
155
Technical Data
Rate hysteresis off -10 ppm off
Parameter/Function
Factory Settings Standard
Program
Safe Program
Repetitive rate
hysteresis
— off —
Scan rate hysteresis — off —
Upper tracking rate 130 ppm 130 ppm —
Dynamic AV Delay Low Low —
AV hysteresis off off —
AV Repetitive
Hysteresis
— — —
AV Scan Hysteresis — — —
Sense compensation
— — —
AV safety delay — — —
Atrial blanking
period
56 ms 56 ms —
Magnet effect Auto Auto Auto
Pulse amplitude V 3.6 V 3.6 V 4.8 V
Pulse width V 0.4 ms 0.4 ms 1.0 ms
Sensitivity A
Sensitivity V
0.2 mV
2.5 mV
0.2 mV
2.5 mV
—
2.5 mV
Refractory Period A
Refractory Period V
425 ms
250 ms
425 ms
250 ms
—
300 ms
Atrial Refractory
Period Extension
0 ms 0 ms —
Tachycardia Behavior
Mode conversion off off —
Mode Switching off on —
X-out-of-8 Criterion — 5-out-of 8 —
Z-out-of-8 Criterion — 5-out-of 8 —
Intervention Rate — 160 —
Far-Field Blanking 56 ms 56 ms —
Switch to — VDIR —
Basic Rate during
Mode Switching
— + 10 —
2:1 Lock-in
Protection
off off —
VES lock-in
protection
off off —
Min. PVARP 235 ms 235 ms —
Sensor Threshold — Medium —