User's Manual
154
Technical Data
Parameter/Function
Factory Settings Standard
Program
Safe Program
Intervention Rate — 160 —
Far-Field Blanking 56 ms 56 ms —
Switch to — DDIR —
Basic Rate during
Mode Switching
— + 10 —
2:1 Lock-in
Protection
off off —
VES lock-in
protection
off off —
Min. PVARP 235 ms 235 ms —
Sensor Threshold — — —
Sensor Gain — — —
Autom. sensor gain — — —
Rate Increase — — —
Max. Activity Rate — — —
Rate Decrease — — —
Rate Fading off off off
RF Rate Increase — — —
RF Rate Decrease — — —
Lead Configuration
Pace A/V unipolar unipolar unipolar
Sense A/V unipolar unipolar unipolar
PMT Management off on —
VA Criterion — 380 ms —
Autom. Lead
Monitoring A/V
off off off
Auto-Initialization on — —
ACC off ATM off
Max. Amplitude — 3.6 V —
Scan Time — at intervals —
Interval — 12 hours —
Cylos VR
Parameter/Function
Factory Settings Standard
Program
Safe Program
Mode VDD VDD VVI
Basic Rate 60 ppm 60 ppm 70 ppm
Night Program off off off