User's Manual

153
Technical Data
Default Programs
Cylos DR/DR-T
Parameter/Function
Factory Settings Standard
Program
Safe Program
Mode DDD DDD VVI
Basic rate 60 ppm 60 ppm 70 ppm
Night program Off Off Off
Rate hysteresis Off Off Off
Repetitive rate
hysteresis
Scan rate hysteresis
Upper tracking rate 130 ppm 130 ppm
Dynamic AV delay Low Low
AV hysteresis off off
Repetitive AV
Hysteresis
AV Scan Hysteresis
Sense compensation
-45 ms -45 ms
AV safety delay 100 ms 100 ms
Atrial blanking
period
56 ms 56 ms
Ventr. blanking
period
32 ms 32 ms
Magnet effect Asynchronous Auto Auto
Pulse amplitude A
Pulse amplitude V
3.6 V
3.6 V
3.6 V
3.6 V
4.8 V
Pulse width A
Pulse width V
0.4 ms
0.4 ms
0.4 ms
0.4 ms
1.0 ms
Sensitivity A
Sensitivity V
1.0 mV
2.5 mV
1.0 mV
2.5 mV
2.5 mV
Refractory period A
Refractory period V
425 ms
250 ms
425 ms
250 ms
300 ms
Atr. refractory per.
ext.
0 ms 0 ms
Mode conversion off off
Mode Switching off on
X-out-of-8 Criterion 5-out-of 8
Z-out-of-8 Criterion 5-out-of 8