User's Manual
151
Technical Data
Parameters at Replacement Indication
Basic Rate Programmed value minus 11% (minus 4.5 - 11% in
modes DVI(R), DDI(R), DVT(R), and DDI/T(R),
depending on the programmed AV delay)
Magnet Rate
80 ppm for 10 cycles directly after magnet application
(not so in synchronous magnet mode)
Pulse Width Programmed values
Pulse amplitudes – programmed values with ACC deactivated (Off)
– last measured threshold + 1.2 V with ACC activated
(On) before ERI was reached
Sensitivity Programmed values