User's Manual
148
Technical Data
Sensitivity A
Sensitivity V
0.1 ... (0.1) ... 1.5 ... (0.5) ... 7.5 mV
0.5...(0.5)...7.5 mV
Cylos VR
Refractory period A
Refractory period V
200 ... (25) ... 775 ms
170; 195; 220; 250; 300; 350; 400 ms
Atrial Refractory
Period Extension
0 ... (50) ... 350 ms
Tachycardia
Behavior
Off, Mode Conversion, Mode Switching
Mode conversion Off; On (in mode VDD(R))
Mode Switching Off; On (in mode VDD(R))
Intervention Rate 180...(10)...250 ppm
X-out-of-8 Criterion 3...(1)...8
Z-out-of-8 Criterion 3...(1)...8
Basic Rate during
Mode Switching
+5...(5)...+30 ppm
2:1 Lock-in
Protection
On; Off
PMT Management Off; On
VA Criterion 250...(10)...500 ms
Min. PVARP Off; On
Lead configuration
for
V pacing
A sensing
V sensing
unipolar; bipolar
bipolar
unipolar; bipolar
Automatic Lead
Monitoring
Off; On
Active capture
control (ACC)
On; Off; ATM
Minimum Ventricular
Amplitude
0.2...(0.1)...3.6...(0.1)...4.8 V
Maximum
Ventricular
Amplitude
2.4; 3.6; 4.8; 6.4 V
Scan Period Intervals; Times of Day
Intervals every 0.1; 0.3; 1; 3; 6; 12; 24 hours
Times of Day,
the 1
st
/2
nd
Time of
Day
0:00 to 24:00 hours
Safety Margin 0.3...(0.1)...(0.5) ...(0.1)...1.2 V
Auto-Initialization Off; Lead Detection; On
VES Lock-in
Protection
On; Off
Termination after 4; 6; 12 cycles
a) The corresponding intervals (t) correlate with the rates (r) according to the formula
t = 60,000 / r (where t is in ms, r in ppm)