User's Manual

147
Technical Data
Cylos DR/DR-T
Scan Period Intervals; Times of Day
Intervals every 0.1; 0.3; 1; 3; 6; 12; 24 hours
Times of Day,
the 1
st
/2
nd
Time of
Day
0:00 to 24:00 hours
Safety Margin 0.3...(0.1)...(0.5) ...(0.1)...1.2 V
Auto-Initialization Off; Lead Detection; On
VES Lock-in
Protection
On; Off
Termination after 4; 6; 12 cycles
a) The corresponding intervals (t) correlate with the rates (r) according to the formula
t = 60,000 / r (where t is in ms, r in ppm)
b) The values 30, 31, 32, 33, and 34 are for temporary settings only.
c) In the event of an electronic defect
d) The values depend on the atrial blanking period set
Cylos VR
Basic rate
a), b)
30 ... (1) ... 88 ... (2) ... 122 ... (3) ... 140 ... (5) ...
180 ppm
Night Program Off; On
Rate Hysteresis
a)
off; -5 ... (5) ... -80 bpm
Repetitive rate
hysteresis
off; 1 ... (1) ... 10
Scan rate hysteresis off; 1 ... (1) ... 10
Upper Tracking Rate
(UTR)
a)
100; 110; 120; 130; 140; 160; 185 ppm
Tachycardia Mode 2:1; WRL (automatic adjustment)
Runaway
protection
a), c)
195 ... 220 ppm
Dynamic AV Delay low; medium; high; individual; fixed
AV Delay Values 15; 50; 75; 100; 120 ... (10) ... 200; 225; 250; 300
ms (programmable in 5 ranges)
AV hysteresis Off; low; medium; high; negative
AV Repetitive
Hysteresis
Off; 1...(1)...6
AV Scan Hysteresis Off; 1...(1)...6
Negative AV
Hysteresis
1...(1)...10...(3)...100...(10)...180
Magnet Effect auto; asynchronous; synchronous
Atrial Blanking
Period
32; 40; 48; 56; 72 ms
Far- Field Blanking 56...(25)...200 ms
Pulse Amplitude V 0.1 ... (0.1) ... 4.8 ... (0.2) ... 8.4 V
Pulse Width V 0.1; 0.2; 0.3; 0.4; 0.5; 0.75; 1.0; 1.5 ms