User's Manual

144
Technical Data
Active Capture Control (ACC)
Status Amplitude
Control (ACC)
On; Off; Deactivated
Ventricular
Thresholds
<
0.3; 0.3; 0.5;... (0.2) ... 4.7;
4.8
Pulse and Timing Parameters
1
Cylos DR/DR-T
Basic Rate
a), b)
30 ... (1) ... 88 ... (2) ... 122 ... (3) ... 140 ... (5) ...
180 ppm
Night Program off; on
Rate Hysteresis
a)
off; -5 ... (5) ... -50 bpm
Repetitive rate
hysteresis
off; 1 ... (1) ... 10
Scan rate hysteresis off; 1 ... (1) ... 10
Upper Tracking Rate
(UTR)
a)
100; 110; 120; 130; 140; 160; 185 ppm
Tachycardia Mode 2:1; WRL (automatic adjustment)
Runaway
protection
a), c)
195 ... 220 ppm
Dynamic AV Delay Off, low; medium; high; individual; fixed
AV Delay Values 15; 50; 75; 100; 120 ... (10) ... 200; 225; 250; 300
ms (programmable in 5 ranges)
AV hysteresis Off; low; medium; high; negative
AV Repetitive
Hysteresis
Off; 1...(1)...6
AV Scan Hysteresis Off; 1...(1)...6
Repetitive Negative
AV Hysteresis
1...(1)...10...(3)...100...(10)...180
Sense Compensation
Off, -15 ... (-15) ... -120 ms
AV Safety Delay 100 ms
Atrial Blanking
Period
32; 40; 48; 56; 72 ms
Far- Field Blanking 56...(25)...200 ms
Ventricular blanking
period
d)
16; 24; 32; 40; 48; 56; 72 ms
Magnet Effect auto; asynchronous; synchronous
1
37°C, 500