User's Manual
143
Technical Data
Atrial Rhythms
Mean Value AES 0; 1...(10);
>
10;
>
100
Number of Atrial
Tachycardias (AT)
0; 1...(10);
>
10;
>
20
Number of Atrial
Fibrillation (AF)
0; 1...(10);
>
10;
>
20
Number of Atrial
Flutter (AFl)
0; 1...(10);
>
10;
>
20
AV Synchrony
(Ax Vx/Vx)
0; 3...(3)...100%
Number of
Tachycardia
Episodes
0 ... (1)...10...(2)...60; > 60
Duration of
Tachycardia
Episodes
0; 3...(3) ...100%
Ventricular Rhythms
Mean Ven. Heart
Rate
<
50; 52; ...(2)...174;
>
174 bpm
Max. Ven. Heart
Rate
a)
<
85; 85;)...248;
>
248 bpm
Duration of Max.
Ven. Heart Rate
a)
<
0.5; 0.5 ...1.0; 1.0 ...2.0; 2.0 ...2.5; 5 min
Max. Ven. Heart Rate
during Tachycardia
Episodes
b)
<
120; 120; 140; 160; 180; 200; 220 bpm
Number of
Max. VES/h
0;1...10; 11...30;
>
30
Number of Ven.
Runs
0; 1; 2; (1)...10;
>
10
Number of Ven.
Episodes (VT)
0; 1; 2;
>
2
a) Captured by IEGM recording during high ventricular rates
b) Captured by IEGM recording of Mode Switching
System Status
Atrial/Ven. Lead
Check
OK; not OK
a)
,
(if not OK, then switch from bipolar to unipolar)
Mean P/R-Wave
Amplitude
No measurement
b)
;
<
50%
a)
; 50 -
<
100%; 100% safety margin
(Battery) Status ERI OK; ERI active
a)
a) Parameter value triggers event report
b) Or the measurement is below the programmed sensitivity