User's Manual
142
Technical Data
Technical Data
Pacing Modes
Cylos DR
DDD-CLS, VVI-CLS, DDDR, DDTR/A, DDTR/V, DDTR,
DDIR, DDIR/T, DVIR, DVTR, D00R, VDDR, VDTR,
VDIR, VVIR, VVTR, V00R, AAIR, AATR, A00R
DDD, DDT/A, DDT/V, DDT, DDI, DDI/T, DVI, DVT,
D00, VDD, VDT, VDI, VVI, VVT, V00, AAI, AAT, A00,
OFF
DDD(R)
+
, DDT/A(R)
+
, DDT/V(R)
+
, AAI(R)
+
, AAT(R)
+
Cylos VR VVI-CLS, VVIR, VVTR, V00R, VVI, VVT, V00, OFF
Valid for
Cylos DR-T
Home Monitoring is possible for the following modes:
DDD(R) , DDT(R)/A, DDT(R)/V, DDT(R), DDI(R),
DDI(R)/T, VDD(R), VDT(R), VVI(R),VDI(R), DDD(R)+,
DDT(R)/A+,
DDT(R)/V+; DDD-CLS; VVI-CLS
Home Monitoring — Programmable Parameters
Valid for Cylos DR-T
Home Monitoring Off, On
Monitoring Interval 1 day
Time of Trend Message
Between 0:00 (12:00 a.m.)
...(10)...
and 23:50 (11:50 p.m.)
Patient Message Off, On
Event Message Off, On
Home Monitoring –
Non-Programmable Parameters/Value Ranges
Valid for
Cylos DR-T
For Home Monitoring, stored data and events are displayed under the
following topics:
— Stored Messages
— Atrial Rhythms
— Ventricular Rhythms
— System Status
Stored Messages
The following are displayed:
— Type and time of last message
— Elapsed time in days
— Number of trend messages
— Number of patient messages