User's Manual
132
Replacement Indication
Remaining Service Time after ERI
The following tables show the mean
1
and minimum
2
values for the
remaining service time between reaching the ERI (elective replacement
indication) and magnet rate after reaching replacement indication EOS
(end of service) for the standard program
3
and a program with a higher
pulse energy.
4
The data are based on a lead impedance of 500 Ohms, 100% pacing
and the data supplied by the battery manufacturer. These times are at
most 30% shorter at a lead impedance of 300 Ohms instead of 500
Ohms.
Cylos DR/DR-T Expected times in DDDR mode
(in months)
Standard
Program
Program with
Higher Pulse
Energy
Mean Value 9 8 Beginning of
ERI to EOS
Minimum Value 7 6
Table 12: Expected service times for Cylos DR
Cylos VR Expected times in VDDR mode
(in months)
Standard
Program
Program with
Higher Pulse
Energy
Mean Value 8 8 Beginning of
ERI to EOS
Minimum Value 6 6
Table 13: Expected service times for Cylos VR
Note: The expected service times could differ from those
given here if program settings are different from
those listed in the above tables.
1
50% of the pacemakers reach or exceed these values
2
99.9% of the pacemakers reach or exceed these values
3
Pulse amp. A/V 3.6 V, pulse width A/V 0.4 ms, rate 60 ppm
4
Pulse amp. A/V 4.8 V, pulse width A/V 1.0 ms, rate 90 ppm