Handling and Implantation
Connecting Cylos DR/DR-T with an IS-1 Connector
Insert the lead connector into the connector receptacle without
bending the lead until the connector pin becomes visible behind the set
screw block.
A: Using the screwdriver included, pierce the slot of the silicone plug
vertically and insert the blade of the screwdriver into the setscrew.
B: Tighten setscrew with the enclosed screwdriver clockwise until the
torque becomes limited (you will hear a crackling sound). Carefully
withdraw the screwdriver without turning back the setscrew.
When you withdraw the screwdriver, the silicone plug automatically
seals the lead connector block safely. The proximal pole of the bipolar
lead is automatically connected. Now attach the second lead connector
as described above.
Insert the non-absorbable fixation suture through the opening in the
lead connector block and fixate the pacemaker in the prepared pocket.