User's Manual

113
Follow-up Options
Note: Program the magnet effect to [synchronous] when
you want the patient to do IEGM recording.
Caution! Due to the compression and reconstruction
processes that the signals undergo, the IEGM
recordings are not suitable for direct morphologic
analyses. If you have activated the "patient-
triggered IEGM recording" function, please tell the
patient how to use the magnet to trigger an IEGM
recording.
Have the patient review the information included
with the pacemaker, including the section entitled
"Storing Intracardiac Data Through Magnet
Application."
Analog Telemetry of Battery, Pulse and Lead Data
The following pulse, battery, and lead data can be measured non-
invasively by means of analog telemetry:
Parameters Unit of Measurement
Battery Voltage V
Battery Impedance k
z
Battery Current µA
Pulse Voltage V
Pulse Current mA
Pulse Energy µJ
Pulse Charge µC
Lead Impedance
z
Table 9: Measurable parameters of analog telemetry