User Manual
19 Guidelines
9Technical Data
General • Operating mode: Continuous operation
•Longevity: 6years
• IP 21: Protection against dripping water
Permissible ambient
conditions
• Operating temperature: -5°C to +40°C
• Storage and transport temperature: -20°C to +60°C
• Store in a dry place:
— Relative humidity: 30% to 75% (non-condensing)
• Atmospheric pressure: 700 hPa to 1060 hPa
CardioMessenger • Type: CardioMessenger Smart
• Dimensions (WxLxH): approx. 67 x 135 x 17 mm
• Weight: approx. 110 g
• MICS frequencies: 402–405 MHz, FSK modulation
• MICS transmission power: 25 µW EIRP
• GSM frequencies: 850 MHz, 900 MHz, 1800 MHz, 1900 MHz
UMTS frequencies: WCDMA band (I, II, IV, V, VI, VIII, and XIX)
• GSM transmission power: 2 watts (850/900 MHz); 1 watt (1800/1900 MHz)
UMTS transmission power: 0.25 W
Power supply brick • Type: FRIWO FW752O/05
• Dimensions (WxHxD): approx. 51.5 x 87.5 x 34 mm
• Housing type: MMP 15
• Housing material: PC / ABS VO 125
• Input voltage: 100–240 V AC at 50–60 Hz
•Output voltage: 5VDC; 3A
• Power cord: approx. 1.83 m
• Operating temperature: 0°C to +40°C
• Storage and transport temperature: -40 °C to +70 °C
• Store in a dry place:
— Relative humidity: 30% to 75% (non-condensing)
• Atmospheric pressure: 700 hPa to 1060 hPa
Battery (integrated) • Type: Lithium-ion