13 Error Resolution
Error type B
Self-test failed The information icon is displayed.
Repeat the self-test because the CardioMessenger did not finish it.
The CardioMessenger is not ready for use.
The CardioMessenger starts and automatically repeats the self-test. After the self-
test has been completed, the CardioMessenger is ready for use.
If the information icon continues to flash, the CardioMessenger is defective. Return
it to your physician.
Error type C
No mobile connection The information icon flashes and the battery icon is displayed. Check the mobile
connection because the CardioMessenger cannot connect to the BIOTRONIK
Service Center.
The CardioMessenger restarts and performs the self-test. It checks the connection
to the cellular phone network. The operation and battery icon is displayed once the
connection test is completed successfully. The CardioMessenger is now ready for
use.
If the CardioMessenger generally cannot connect to the cellular phone network
from positions near your bed, contact your physician.
1 Press the on/off key on the right of the CardioMessenger for approxi-
mately five seconds.
2 Press the on/off key again for approximately five seconds.
Note: The connection test can take up to 15 minutes.
1 Press the on/off key on the right of the CardioMessenger for approxi-
mately five seconds.
2 Put the CardioMessenger in a place with better mobile reception.
Make sure that the distance to the device is still less than 2 m (6 feet).
3 Press the on/off key again for approximately five seconds.