Instruction manual
55 
responsive to ethanol than to methane, so the readings 
will be low. Multiplying the instrument reading by the 
correction factor for ethanol will produce the true 
concentration. 
Given that the correction factor for ethanol is 1.2, if the 
instrument reading is 40 percent LEL, then the true 
concentration is seen to be about 48% LEL. 
(40 % LEL) X (1.2) = 48% LEL) 
Instrument Correction Actual 
 Reading Factor Concentration 
It is important to note that the correction factor for ethanol 
is different when the instrument is calibrated on propane. 
In the case of a propane calibrated instrument, 
instrument readings for ethanol will be high. Given that 
the correction factor for ethanol in this case is 0.76; when 
the instrument reads 40 percent LEL, the true 
concentration for ethanol is 30% LEL. 
(40 % LEL) X (.76) = (30% LEL) 
Instrument Correction Actual 
 Reading Factor Concentration 
2. Effects of contaminants on combustible sensors 
Combustible sensors may be affected by exposure to 
silicone containing substances (found in many lubricants 
and hydraulic fluids), the tetra-ethyl-lead in "leaded" 
gasoline, and halogenated hydrocarbons (Freons
®
, or 
solvents such as trichloroethylene and methylene 
chloride). High concentrations of hydrogen sulfide may 
also damage the sensor. 
If sensitivity of the combustible sensor is lost due to 
poisoning, it tends to be lost first with regards to 
methane!  
A partially poisoned sensor might still respond accurately 
to propane while showing a dangerously reduced 
response to methane. 
2.3. Choosing the right calibration gas for 
combustible sensors 
The best results are obtained when calibration is done 
using the same gas that is expected to be encountered 
while actually using the instrument. When not sure what 
combustible gases might be encountered, it is important 
to choose a calibration gas that will provide a level of 
sensitivity which is typical of the widest range of 
combustible gases. 
Propane provides a sensor response which is more 
typical of the wide range of combustible gases and 
vapors than any other calibration mixture. The only 
drawback to using propane based calibration gas 
mixtures is that a partially poisoned sensor might still 
respond accurately to propane while showing a 
dangerously reduced response to methane. 
Use of Biosystems’ “Propane Equivalent” calibration 
mixtures guards against this potentially dangerous sort of 
calibration error. 
Biosystems’ “Propane Equivalent” calibration mixtures 
are based on methane, but in concentrations which are 
designed to produce a level of sensitivity “equivalent” to 
that provided by a mixture which contains a 50% LEL 
concentration of propane. Because Biosystems’ 
equivalent mixtures are based on methane, any loss of 
sensitivity to methane is detected (and can be corrected) 
immediately. 
Use of other gases such as pentane or hexane to 
calibrate the instrument should be reserved for situations 
where these are either the gases predominantly present, 
or where the relative response to the calibration gas 
closely approximates that of the actual gas to be 
measured. 
Verifying accuracy before each day’s use insures that 
proper sensitivity is maintained over the life of the 
combustible sensor. 
3. Toxic gas alarms 
The PhD Ultra has three separate alarm points for toxic 
gases: Ceiling, STEL, and TWA. 
OSHA has assigned some, but not all, toxic substances 
with a ceiling or "Peak" exposure level. This is the 
highest concentration of a toxic substance to which an 
unprotected worker should ever be exposed, even for a 
very short time. Never enter an environment even 
momentarily when concentrations of toxic substances 
exceed the ceiling level. 
The Time Weighted Average ( or TWA) is the maximum 
average concentration to which an unprotected worker 
may be exposed over an eight hour working day. During 
this time, STEL and ceiling concentration limits may not 
be exceeded. 
OSHA has assigned some, but not all, toxic substances 
with a Short Term Exposure Limit. The STEL is the 
maximum average concentration to which an unprotected 
worker may be exposed in any fifteen minute interval 
during the day. During this time, neither the eight hour 
TWA or the ceiling concentration may be exceeded. Any 
fifteen minute periods in which the average STEL 
concentration exceeds the permissible eight hour TWA 
must be separated from each other by at least one hour. 
A maximum of four of these periods are allowed per eight 
hour shift. 
The table below shows the highest levels at which these 
alarms should be set. If OSHA has not determined a 
ceiling value, for greatest safety the PhD Ultra ceiling 
alarm should be set at the same value as the STEL 
alarm. If OSHA has not determined a STEL value, the 
PhD Ultra STEL alarm should be set at the same value 
as the TWA. 
4. U. S. Department of Labor, Occupational Safety 
and Health Administration (OSHA) Permissible 
Exposure Limits for select toxic gases: 










