Instruction manual
50 
LCD display indicates that the pump is attached and 
in normal operation.) 
The sample draw pump includes a unique pressure 
sensor designed to protect the PhD Ultra from 
exposure to water or other liquids. If there is a 
change in pressure in the sample draw assembly 
due to fluid intake, the pump immediately shuts 
down. After a few seconds audible and visual 
alarms indicating a low flow condition will also be 
activated. 
CAUTION:  Never perform remote sampling 
with the PhD Ultra without the sample probe 
assembly. The sample probe handle contains 
replaceable filters designed to block moisture 
and remove particulate contaminants. If the 
pump is operated without the probe assembly in 
place, contaminants may cause damage to the 
pump, sensors and internal components of the 
PhD Ultra. 
Procedures for proper use of the motorized 
sample pump are contained in Chapter 2. 
Replacement of sample probe filters is 
discussed in Section 2.4.3.1. 
As an additional safeguard, the pump also contains 
an internally housed particulate filter. If the pump is 
operated without the sample probe assembly in 
place, or in particularly dirty atmosphere, this 
internal filter can become clogged and periodically 
require replacement. Standard accessories 
included with every 54-05-A0101 motorized pump 
include a package of 10 replacement 61-001 filters. 
5.4.1  Internal pump filter replacement 
Use the following procedure for replacing the 
internal pump filter: 
(1) Remove the three bottom screws from the pump 
assembly and disassemble the upper and lower 
sections of the pump case as shown in Figure 
5.4.3. 
(2) Remove the two pieces of tubing from the filter 
cap by gently twisting and pulling until the ends 
are free. 
(3) Remove the filter cap and replace the used 61-
001 filter. 
(4) Reattach the filter cap and tubing. Make sure 
the tubing is secure to the filter cap before 
reassembling the pump case. 
(5) Reassemble the upper and lower sections of the 
pump case. Do not over-tighten the three 
screws! 
Proper pump operation must be verified before 
the pump is put back into service. 
Use the following procedure to verify pump 
performance: 
(1) Attach the pump to your PhD Ultra and turn the 
instrument on. Wait for the instrument to 
complete the self-test sequence. 
(2) Verify that the pump is operating normally, and 
that a flashing “P” indicating proper pump 
performance may be seen in the upper left hand 
corner of the instrument LCD. 
(3) Remove the hose and probe assembly from the 
pump (if currently attached) then cover the 
pump inlet with a finger. If there are no leaks in 
the internal pneumatics the pump should go into 
a low-flow alarm and shut down, and the audible 
and visual low flow alarms should be activated. 
A message screen will identify that there is a 
low pump flow condition. A second screen will 
advise you to remove the blockage and press 
“mode” to resume operation. 
(4) Press “mode” to reset the pump and continue 
the test. 
(5) Reattach the hose and probe assembly. Cover 
the end of the sample probe tube, and verify 
once again that the pump shuts down and that 
the low flow alarms are activated. 
(6) Press “mode” to reset the pump and resume 
normal operation. 
Figure 5.4.1. Motorized sample draw pump and 
probe assembly 
Figure 5.4.2. Top and bottom views of motorized 
54-05-A0101 sample pump case 










